6. INTERFACES
PC
T e s t e r
9pol 25pol (female) (female) 25pol 9pol
5
7
—GND—————————— GND—— 7 5
2
3
—RxD——— ————— RxD—— 3 2
3
2
—TxD———- ———— TxD—— 2 3
7
4
—RTS———- ———— RTS—— 4 7
8
5
—CTS———- ———— CTS—— 5 8
4
20 —DTR——— ———— DTR—— 20 4
6
6
—DSR—o—-- ——o—- DSR—— 6 6
1
8
—DCD— —
DCD—— 8 1
6.2
Interfaces
6.2.1 RS232/USB/Ethernet interface X9
The serial RS232 interface is installed in the device as the standard inter-
face. The 9-pin D-SUB socket X9 is provided for the connection to a control
computer. The connections DTR, DSR, DCD are not essential necessary.
The interface and the control commands are described in the separate
manual „Remote control via RS232/Ethernet".
Fig. 6.2: Interface cable
X9
also as USB-INTERFACE or Ethernet-INTERFACE possible.
For device support: Driver-software and driver-manuals are located at the
ELABO Support CD, enclosed to the device manual.
6.2.2 USB-Interface X8/ Ethernet-Interface X9a
X8
USB-interface for software-update at tester with touch display - see
chapter. 3.15.
X9a
Ethernet-interface for controlling external devices, transformers,
switching units ...