96
11. Test parameter
Full-electronic HV-Tester F1-1x Typs
11.4 Setting parameters function test (FK)
On this menu the settings for the
function test are made. The option
F9-7M E20 (function test) is required for the function test
.
Function tests are undertaken at function voltage, this can be a voltage
from an internal or external voltage source or also at mains voltage. During
the function test, current, voltage and power are measured and evaluated.
Depending on which device variant you have, some of the parameters
given in the following may not be available or it may not be possible edit
some parameters. The values entered are changed in the parameter set
current open. The following parameters relate to a function plug-in with
integrated AC source 0 ... 265 V / 16 A output.
11.4.1
Voltage
Test voltage (V)
- defines the voltage for the function test. The integrated
AC source regulates the voltage, the integrated switch panel switches
the voltage to the sockets.
Minimum voltage „on/off“
- activates the voltage monitoring for the
test object supply. The test voltage must not drop below the defined
minimum voltage during the test time. If this is the case, the device
indicates U>Umin.
off
- no minimum voltage monitoring.
on
- minimum voltage monitoring active.
Minimum voltage (V)
- the voltage must not drop below the pre-defined
minimum voltage during the function test time.
Fig. 11.14: Function test parameters - Voltage