79
10. Menu System
Full-electronic HV-Tester F1-1x Typs
10.9.6
Special
Using the „Special“ tab you can make settings that are only available or
relevant for this device variant. Critical parameters are again password
protected.
Reset safety current limitation
(only on devices with current limitation):
A fault in the current limitation can be reset. This fault occurs only
rarely in unfavourable circumstances for technical reasons.
Important
: The current limitation is only allowed to be reset once.
If the fault occurs again, a fault in the monitoring is to be as-
sumed. For safety reasons the device must be sent to the
manufacturer for checking.
„
Reset safety current limitation
“ is protected using a special
password. The password can be requested from ELABO.
Fig. 10.27: Setting device parameters - Special