68
10. Menu System
Full-electronic HV-Tester F1-1x Typs
Available steps
- this column indicates the existing test steps for the
function (test type) selected. General test steps do not generally have
test steps.
„Test steps“
- this column displays the test steps and sequence in the
test plan.
1
3
4
2
Fig. 10.17: Test plan management - assembling test plan
Assemble test plans as follows
Select the kind of test - HV, IS, PE, Pause, etc.
The column „Existing steps" displays all existing (already defined) steps
for the selected kind of test.
Define the position in the test plan for the new test step. For this pur-
pose, mark in the test plan the test step in front of which the new step
is to be added, e.g. IS500
.
Select the required step, e.g. „Pause5s“.
Touch the „
>>
“ button.
The selected test step is inserted in the list in front of the step marked.
„New“
- opens the parameter screen for the selected test type, e.g. HV.
Adjust the test parameters and save the new test step by using a new
name. See also chapter 10.5. „Load parameter“.
„>>“
- inserts the test step marked in the list (test plan).
"Remove"
- deletes the test step marked from the list (after prompt).
„Change“
- for a temporary modifying of the test parameter. Opens the
parameter screen for the selected test step, e.g. HV200DC. Adjust the
test parameters and save the test step. Overwrite the old parameter of
the step or create a new step by renaming.