81
11. Test parameter
Full-electronic HV-Tester F1-1x Typs
11
TEST PARAMETERS
11.1
Setting parameters high voltage test
On this menu the settings for the HV tests are made. Depending on which
device variant you have, some of the parameters given in the following may
not be available or it may not be possible edit some parameters. The values
entered are changed in the parameter set current open. The parameters
in the following relate to the standard device with 2500 V/5000 V/ 100 mA.
In case of devices with options fitted, the parameters may vary. On this
topic see chapter 14 Options.
11.1.1 TEST VOLTAGE
Voltage „AC / DC (option)“
AC
- HV testing using 2500/5000 V AC max.100 mA (example).
DC
(option) - HV testing with 3000/6000 V DC max.10 mA.
Voltage range „2.5 / 5.0 kV AC and 3.0 / 6.0 kV DC (option)“
2.5 AC / 3.0 DC
(option) - low voltage range.
5.0 AC / 6.0 DC
(option) - high voltage range.
Test voltage „2.5 / 5.0 kV“
- enter the test voltage with which the HV
testing is to be undertaken. For example:
low
voltage range
0,1 ... 2.5 kV AC / 0.2 ... 3,0 kV DC
(option)
high
voltage range
0.2... 5.0 kV AC / 0.3 ... 6.0 kV DC
(option)
Start up voltage „on/off
“ - defines whether the test voltage is switched
on at the lowest possible voltage or from a higher pre-defined voltage.
Fig. 11.1: HV test parameters - Test voltage