Instructions for use Fabius plus SW 3.n
97
Operation
The following alarms can be activated or
deactivated on the
ManSpont
screen with the
ON
/
OFF
key (B):
–
Apnoea Pressure
(see chapter ''Pressure
alarms in automatic ventilation modes'' on
page 128)
–
Volume Alarms
(see chapter ''Volume alarms''
on page 127)
Changing to manual ventilation
1
Set the APL valve head to the desired
maximum airway pressure.
Settings between the grid marks are also
possible.
2
Press the
ManSpont
(A) key.
3
Confirm the new mode.
25
466
ManSpont
B
B
NOTE
In
ManSpont
mode, the apnea alarm time for
triggering apnea volume alarms is extended from
15 seconds to 30 seconds (Caution category)
and from 30 seconds to 60 seconds (Warning
category).