Operation
104
Instructions for use Fabius plus SW 3.n
In the following table, all parameters (B) in
Pressure Support
mode, together with the
corresponding setting ranges and factory settings,
are listed.
If the inspiratory flow (
Insp Flow
) during
inspiratory effort is greater than the set trigger flow
(
Trigger
), the device supports the patient with the
setting
∆
PPS
.
The set inspiratory flow (
Insp Flow
) defines how
fast the
∆
PPS
pressure is reached. When 25 % of
the maximum inspiratory flow (
Insp Flow
) is
reached (or after maximum 4 seconds), the
inspiration is automatically ended. The value
Freq
Min
(e.g., 3 bpm (1/min)) defines a safety period
(safety period = 1/
Freq Min
, e.g., 20 seconds). If
no inspiratory effort is detected and the safety
period has elapsed, the device generates a
pressure-controlled breath with
PINSP
=
∆
PPS
.
The lower alarm limit of the airway pressure
PAW
low
is used for the airway pressure monitoring to
detect apnea (disconnection) and continuous
pressure.
If the pressure waveform does not cross the
pressure threshold either from above or below, an
alarm is sounded.
Parameter
Setting range Factory set-
ting
Support pres-
sure
∆
PPS
[cmH
2
O]
([hPa])
3 to 20,
OFF
10
Minimum respi-
ratory rate for
apnea ventila-
tion
Freq Min
[bpm] ([1/min])
3 to 20,
OFF
3
Trigger sensi-
tivity
Trigger
[L/min]
2 to 15
2
Inspiratory flow
Insp Flow
[L/min]
10 to 85
30
PEEP
[cmH
2
O]
([hPa])
0 to 20
0
21743
PD[
VHF
$SQHD
9HQWLODWLRQ
7LPH
7LPH
)ORZWULJJHU
)ORZWULJJHU
QRIORZWULJJHU
)ORZ
ORZ
PAW
Trigger
Trigger
∆
PPS
Insp Flow
Freq Min
Freq Min
PAW