115
Operation
Preparing for storage or transport
1
Press the
button (A) and confirm.
Monitoring and alarm monitoring are switched
off. The ventilator stops.
2
Set the control dial (B) of the vaporizer until it
engages on
0
.
3
Close the flow control valves.
4
Close the cylinder valves of the gas cylinders.
5
Pull off the O
2
sensor from the inspiratory valve
and expose to the ambient air.
6
Switch off Fabius with the On/Off switch (B) on
the rear and pull out the plug.
7
Remove hoses of the anesthetic gas receiving
system.
WARNING
Risk of tipping over during transport
The medical device may tip over if handled
incorrectly. Observe the following points
when transporting medical devices:
– The medical device may only be moved by
people who have the physical ability to do
so.
– To improve the maneuverability, transport
the device with 2 persons.
– When transporting over inclines, around
corners, or over thresholds (e.g., through
doors or in elevators), make sure that the
medical device does not bump against
anything.
– Remove any devices mounted to the
holding arms or the top of the device.
– Clear the writing tray and fold it down
completely or slide it into the device.
– Do not pull the medical device over hoses,
cables, or other obstacles lying on the
floor.
– Do not activate the brake while the
medical device is being moved.
– Always use the handles on the device to
push or pull it.