Cleaning, disinfection and sterilization
172
Instructions for use Fabius plus SW 3.n
Removing the compact breathing system
Before removing the compact breathing system,
the following hoses and cables must be removed:
– Flow sensor cable
– O
2
sensor cable and O
2
sensor capsule
– Pressure measurement hose
– APL bypass hose
– PEEP/PMAX hose
Removing the inspiratory valve
1
Remove the valve cover plug (B) or the O
2
sensor from the dome (C) of the inspiratory
valve.
2
Screw off the cap nut (A).
3
Remove the dome (C).
4
Take out the valve plate (D).
5
Remove the sealing ring (E) from the socket
(F).
Removing the expiratory valve
1
Screw off the cap nut (G).
2
Remove the sight glass (H).
3
Take out the valve plate (I).
4
Remove the sealing ring (J) from the socket
(K).
Removing the exhaust port
Screw off the exhaust port (A).
20980
A
B
C
D
E
F
G
H
I
J
K
20994
A