Assembly and preparation
58
Instructions for use Fabius plus SW 3.n
Gas cylinders with screw connections
(optional)
Have service personnel repair any leaky or stiff
gas cylinder valves.
1
Place the full gas cylinders (A) in the cylinder
holder and secure them.
2
Connect the pressure reducers to the cylinder
valves. The connections must fit directly to
each other. Do not use transition pieces.
3
Screw the compressed gas hoses (B) to the
pressure reducers and to the connections of
the gas supply block.
4
Open the cylinder valves.
WARNING
Risk of explosion
When pressurized, O
2
is self-igniting in
combination with oil or grease.
Do not oil or grease the gas cylinder valve or
the pressure reducer of the O
2
cylinder. Do
not touch with oily or greasy fingers.
WARNING
Risk of explosion
If the gas cylinder valves are opened too
quickly, a sudden increase in pressure may
occur.
– Open and close the gas cylinders valve
slowly by hand.
– Do not use tools.
CAUTION
Risk due to gas supply failure
Leave gas cylinders as a backup at the
anesthesia workstation, even if there is a
connection to the central gas supply.
CAUTION
Risk of patient injury
Do not connect gas cylinders without the
pressure reducers listed in the current list of
accessories.
NOTE
Keep gas cylinders closed as long as they are
not being used. There is the risk of accidental
emptying of the gas cylinders.
2
5417
WARNING
O
2
gas cylinders must not be mounted on the
right side of the device (looking from the
rear).
O2
N2O
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
AIR
O2
N2O
AIR
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
A
B