Instructions for use Fabius plus SW 3.n
181
Cleaning, disinfection and sterilization
Before using on patients again
1
Assemble the device components, see
''Assembly and preparation'' on page 50.
2
Mount the parts in the reverse oder of the
disassembly, see ''Disassembly'' on page 170.
3
Check readiness for operation, see ''Checking
the readiness for operation'' on page 88.
Flow sensor
Weekly
Observe the associated instructions for use.
Items which can be
reprocessed
Recommended
reprocessing
intervals
Pre-
cleaning
Machine
cleaning and
disinfection
Manual
Steriliza-
tion
Cleaning
Disinfec-
tion