Instructions for use Fabius plus SW 3.n
69
Assembly and preparation
Mounting the rigid arm
Prerequisite: Before mounting the rigid breathing
bag holder, the bag elbow on the compact
breathing system must be removed.
1
Position and align the connection adapter (A)
of the rigid arm on the connection of the
compact breathing system.
2
Tighten the knurled screws (B). Check that the
arm is fixed securely.
3
Fasten the breathing bag (C) at the other end
of the rigid arm.
Mounting the bag elbow
1
Attach the bag elbow (A) on the connection of
the compact breathing system.
2
Attach the breathing bag (B) on the connector
of the bag elbow.
21
012
A
B
C
21
013
A
B