
Troubleshooting
164
Instructions for use Fabius plus SW 3.n
(23/31)
APNOEA PRESSURE !!!
The apnea pressure alarm is
based on a time staggering.
In the
Volume Control
,
Pressure Control
,
SIMV/PS
modes with
Freq
≥
6 or in
Pressure Support
mode
with apnea ventilation deacti-
vated:
Warning =
PAW
did not
exceed the pressure thresh-
old value for a duration of >30
seconds.
In the
ManSpont
,
SIMV/PS
modes with
Freq
< 6 or in
Pressure Support
mode
with apnea ventilation acti-
vated:
Warning =
PAW
did not
exceed the pressure thresh-
old value for a duration of >60
seconds
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.
(20)
APNOEA VENTILATION !!
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.
If two or more successive
breaths of the apnea ventila-
tion are automatically trig-
gered, the settings for
Pressure Support
are not
correct.
Fabius detects a sponta-
neous breath of the patient.
Check settings for
Pres-
sure Support
.
(7)
BATTERY LOW !
No mains power and battery
<20 %
Restore mains power sup-
ply.
(17)
BATTERY LOW !!
No mains power and battery
<10 %
Restore mains power sup-
ply.
Alarm
priority
Alarm
Cause
Remedy