Instructions for use Fabius plus SW 3.n
9
For your safety and that of your patients
Accessories
Connected devices
No operation in potentially explosive
areas
WARNING
Risk due to incompatible accessories
Dräger has only tested the compatibility of
accessories that appear in the current list of
accessories or in separate declarations by
Dräger. If other, incompatible accessories are
used, there is a risk of patient injury due to
medical device failure.
Dräger recommends using the medical device
only with accessories from the current list of
accessories.
WARNING
Risk of operating errors and incorrect use
Strictly observe the instructions for use of all
accessory parts, e.g.,:
– Water traps
– Flow sensors
– CLIC adapter
– CLIC absorber
– Soda lime
– Breathing hoses
– Masks
– Filter
– Endotracheal suction
– Vaporizer
– Manual resuscitator
– AGSS terminal unit
WARNING
Risk of electric shock and device malfunction
Any connected devices or device
combinations not complying with the
requirements in these instructions for use
may compromise correct functioning of the
medical device.
Before using the medical device, refer to and
strictly comply with the instructions for use
of all connected devices and device
combinations.
WARNING
Risk of device malfunction
This medical device can be operated in
combination with other Dräger devices or
with devices from other manufacturers. If a
device combination is not approved by
Dräger, the safety and the functional state of
the individual devices can be compromised.
– The operating organization must ensure
that the device combination complies with
the applicable editions of the relevant
standards for medical devices.
– Strictly observe the assembly instructions
and instructions for use of each
connected device.
WARNING
Risk of explosion and fire
This medical device is neither approved nor
certified for use in areas where oxygen
concentrations greater than 25 Vol%,
combustible or explosive gas mixtures are
likely to occur.