Application
18
Instructions for use Fabius plus SW 3.n
Further information on application
Environment of use
Fabius is designed for use in rooms in which
therapeutic or diagnostic interventions can be
carried out.
Do not use Fabius in the following environments:
– Outside of massive buildings
– In intensive care units
– During patient transport
– In vehicles, airplanes, or helicopters
WARNING
Risk of explosion
This medical device is neither approved nor
certified for use in areas where oxygen
concentrations greater than 25 Vol%,
combustible or explosive gas mixtures are
likely to occur.
WARNING
Risk of device malfunctions and/or patient
injury and user injury
Magnetic fields can negatively influence the
correct functioning of the medical device and
therefore endanger the patient or user.
Do not use the medical device near magnetic
resonance imagers.
WARNING
Do not use soda lime based on potassium
hydroxide. Otherwise, there is a risk of CO
formation.