Instructions for use Fabius plus SW 3.n
55
Assembly and preparation
Central gas supply
Version for 3 gases
1
Screw the connection piece of every individual
central gas supply hose hand-tight on the
corresponding connection (A) on the device
side.
2
Plug the probes of the central gas supply
hoses in the appropriate wall terminal units (B).
3
Check if all central gas supply hoses are
correctly connected.
WARNING
Risk to patient through incorrect gas
connection
If the connections of the central gas supply
hoses between the central gas supply and
Fabius are interchanged, serious accidents
can occur.
If Fabius is connected to the central gas
supply, it must be checked whether the marks
on the hoses agree with the connections of
the central gas supply.
25414
O2
N2O
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
AIR
A
B
CAUTION
Risk of insufficient gas pressure
If the supply pressure of the central gas supply is
too low, the functionality of the anesthesia
workstation is compromised.
The pressure gauge of the individual gases must
display a constant pressure between 41 and
87 psi (2.8 and 6 kPa x 100).
NOTE
A failure of the gas supply can lead to a failure of
the connected devices.