Configuration
136
Instructions for use Fabius plus SW 3.n
Results of the system leakage test
The test of the system leakage can have the
following results:
Test of the overpressure safety valve
This test checks the functionality of the
overpressure safety valve.
The test results are displayed on the screen with
the leakage test results (B).
Troubleshooting with failed test of the
overpressure safety valve
Repeat leakage test. If the test of the
overpressure safety valve continues to fail,
contact DrägerService or the authorized local
service partner.
System leakage
[mL/min]
Displayed result
[mL/min]
250
Measured value and
PASSED
251 to 350
Measured value and
FAILED
>350
>350 and
FAILED