Instructions for use Fabius plus SW 3.n
201
Technical data
Vaporizer interface
The anesthesia workstation is equipped with an interlock system.
When removing the vaporizer, the connection is automatically closed and sealed.
The following vaporizers can be used:
– Dräger-Vapor for halothane
– Dräger-Vapor for enflurane
– Dräger-Vapor for isoflurane
– Dräger-Vapor for sevoflurane
– Datex-Ohmeda Devapor/D-Tec for desflurane
– Dräger D-Vapor
Technical data of the vaporizers are contained in the corresponding instructions for use.
Measured value or
waveform
Range
Resolution Accuracy
Condition
PAW
Airway pressure
(numeric)
-20 to 99 cmH
2
O
(hPa)
1 cmH
2
O
(hPa)
±4 %
1)
Airway pressure
(waveform)
0 to 99 cmH
2
O
(hPa)
Pressure gauge
(mechanical)
-20 to 80 cmH
2
O
(hPa)
2 cmH
2
O
(hPa)
1.28 cmH
2
O (hPa)
MV
e
Expiratory minute
volume
0 to 32.0 L/min
0.1 L/min
±15% or ±0.2 mL,
depending on which
value is higher
2)
Based on 20 °C
(68 °F) Ambient
pressure and satu-
rated gas
VT
e
Expiratory tidal
volume
0 to 1500 mL
1 mL
±15 %
2)
or ±20 mL,
depending on which
value is higher