Maintenance
Preventive maintenance
The following table shows the preventive maintenance intervals:
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
WARNING
Risk of electric shock
Before performing any service work,
disconnect all electrical connections and gas
connections from power and gas supplies.
Component
Interval
Measure
Personnel responsible
CO
2
absorber
If colored violet
Replace
Users
Water trap
As needed or if
soiled
Replace
Users
Flow sensor
As needed or if
calibration is no
longer possible
Clean/Replace
Users
Internal lithium battery
Every 36 months
Replace
Experts
Fabius
plus
Every 12 months
Inspection and service
Experts
Breathing system
Every 12 months
Inspection and service
Service personnel
Vaporizers
Every 12 months
Inspection and service
Service personnel
Sensors
Every 12 months
Inspection and service
Service personnel
Lead-gel battery
Every 3 years
Replace
Experts
Cylinder pressure reducer for
high-pressure cylinders
1)
1) optional
After 6 years
Basic overhauling
Experts
Pressure reducer for Pin-
index1)
After 6 years
Replace
Experts