Instructions for use Fabius plus SW 3.n
221
Annex
Fresh-gas delivery and S-ORC
Activate the
ManSpont
mode.
Open the O
2
flow control valve completely.
Check whether the O
2
fresh-gas flow dis-
play shows at least 10 L/min.
Close the Air flow control valve. Open the
N
2
O flow control valve completely. Check
whether the N
2
O fresh-gas flow display
shows at least 10 L/min.
Shut off O
2
supply:
– Remove the O
2
connection of the cen-
tral gas supply.
– Close the O
2
cylinder valve.
– Check whether the red LED for low O
2
supply pressure is flashing.
– N
2
O flow is interrupted.
Restore O
2
supply:
– N
2
O flow is present.
Set the O
2
flow control valve to 1.5 L/min.
N
2
O gas delivery = 3 L/min to 5 L/min.
Close the O
2
flow control valve:
– N
2
O flow is interrupted.
Open the Air flow control valve. Check
whether the Air fresh-gas flow display
shows at least 10 L/min.
Close all flow control valves.
Calibration of the sensors
Remove the O
2
sensor housing from the
cover of the inspiratory valve.
Expose the O
2
sensor to ambient air for
2 minutes.
Start the calibration.
Insert the O
2
sensor housing back in the
cover of the inspiratory valve.
Calibrate the flow sensor.
Check type of gas
Set the O
2
flow control valve to 3 L/min.
Check whether the measured O
2
concen-
tration is at approx. 90 to 100 Vol%.
Close the O
2
flow control valve.
Vapor 19.n, Vapor 2000/3000 (Tec 5)
P
Vaporizer is firmly connected, locked, and
is hanging vertically.
P
Control dial is in position
0
.
P
Filling level is between the minimum mark
and the maximum mark.
P
Safety filling device:
– Sealing block is inserted and closed
tightly.
– Filling inlet is locked.
– Drain valve is closed.
P
Filling device Quik Fil:
– Drain valve is closed.
– Sealing cap is tightly closed.
P
Dräger Fill filling system:
– Drain valve is closed.
– Sealing cap is tightly closed.
D-Vapor (optional)
P
Vaporizer is firmly connected, locked, and
is hanging vertically.
P
Control dial is in position
0
.
P
Filling level is between the minimum mark
and maximum mark.
P
Operation LED is lit.