Instructions for use Fabius plus SW 3.n
131
Configuration
Configuration in standby mode
The following configuration functions are available
in
Standby
mode:
– Calibrations
– System tests
– Management of default settings
1
Press the
(A) key.
The pressure waveform window is replaced by a
confirmation message (B) and instructions to turn
off the flow.
The LED of the
key (A) starts flashing. It
flashes until the
Standby
mode is confirmed.
2
Confirm the new mode.
The ventilator changes to
Standby
mode. The
previous screen is replaced by the start screen
Standby
and the standby LED is continually lit.
The following softkeys (C) are displayed on the
start screen:
2541
1
25491
NOTE
If the confirmation is not done within 15 seconds,
the ventilator remains in the previous mode and
the pressure waveform window is restored.
A
B
ManSpont
25463
–
Run
System
Test
–
Calibrate
Flow
Sensor
–
Calibrate
O
2
Sensor
–
Leak /
Compl
Test
–
Access
Alarm
Log
–
Restore
Site
Defaults
C
Standby