For your safety and that of your patients
Information on electromagnetic
compatibility
General information on electromagnetic
compatibility (EMC) according to the international
EMC standard IEC 60601-1-2:
Medical electric equipment is subject to special
precautionary measures concerning
electromagnetic compatibility (EMC) and must be
installed and put into operation in accordance with
the EMC information provided (see page 190).
Portable and mobile RF communications
equipment can affect medical electrical equipment.
Installing accessories
Stricly observe the instructions for use and
assembly instructions.
Keeping the instructions for use
Training
User training is offered by the responsible Dräger
organization, see www.draeger.com.
WARNING
Risk of electric shock
Do not connect connectors with an
ESD warning symbol and do not
touch the pins of such connectors
without implementing ESD protective
measures. Such protective measures may
include antistatic clothing and shoes,
touching a potential equalization pin before
and during connection of the pins, or using
electrically insulating and antistatic gloves.
All relevant users must be instructed in these
ESD protective measures.
WARNING
Risk of device failure
Electromagnetic fields, e.g., those generated
by radio frequency communication
equipment such as mobile phones, high-
frequency electrical surgery equipment,
defibrillators or shortwave therapy devices
can disrupt the function of the medical
device.
Only operate radio frequency devices at a
sufficient safety clearance of at least 20
cm (7.9 in).
WARNING
Risk of electric shock
The connection of devices to auxiliary power
sockets can lead to an increased leakage
current. If the protective ground of one of
these devices fails, the leakage current may
rise above the permissible values.
– Only connect with the approval of the
respective device manufacturer.
– Have the leakage current checked by
service personnel.
– If the permissible value is exceeded, use a
mains power socket on a wall instead of
the auxiliary power socket of the device.
CAUTION
Risk of device failure
Install the accessory on the basic device in
accordance with the instructions of the basic
device.
Check for secure connection to the basic device.
CAUTION
Risk of incorrect use
The instructions for use must be kept in an
accessible location for users.