Instructions for use Fabius plus SW 3.n
173
Cleaning, disinfection and sterilization
Removing the flow sensor
1
Loosen the expiratory port (A) and remove.
2
Remove the flow-sensor guard (B).
3
Remove the flow sensor (C).
Removing the APL valve
1
Loosen the knurled nut (A).
2
Remove the valve.
Removing the ventilator parts
1
Open the ventilator door.
2
Remove the pressure sensor line (B) of the
ventilator chamber from the corresponding
connection.
3
Release the 3 clamps (D).
4
Remove the cover (A).
5
Remove the ventilator membrane (C).
Removing the anesthetic gas receiving
system
To disassemble, follow the steps in reverse order
as listed in chapter ''Connecting the anesthetic gas
receiving system (optional)'' on page 60.
20982
20945
A
B
C
A
20979
A
B
C
D