Instructions for use Fabius plus SW 3.n
119
Alarms
Alarm priorities
Fabius assigns the appropriate priority to each
alarm.
The background color of the alarm message field
indicates the priority of the active alarm.
The alarm messages are only displayed on
colored background if the option "Color display" is
activated.
The alarm messages are sorted according to these
priorities and displayed corresponding to the
internal priority system. Priority 31 has the highest
and priority 1 the lowest priority. The priority
numbers are listed in the table in chapter ''Alarm –
Cause – Remedy'' on page 160.
A maximum of 4 alarm messages
*
can be
displayed at the same time in a list. Alarm
messages with higher priority are displayed before
alarm messages with lower priority. Alarm
messages with low priority are only displayed if the
cause for a high-priority alarm is remedied.
Color
Priority of the alarm message
Action required
Red
Warning
Alarm with high priority
!!!
Immediate action is necessary in order to
avert imminent danger.
Yellow
Caution
Alarm with medium priority
!!
Fast action is necessary in order to avert a
danger.
Note
Alarm with low priority
!
Attention is necessary, but a delayed
response is sufficient.
*
optionally, 3 alarm messages for Japan and China