Getting started
86
Instructions for use Fabius plus SW 3.n
Daily checkout and pre-use checkout
After preparing the medical device, the daily
checkout and the pre-use checkout must be
performed according to the appendix to these
instructions for use. This ensures that the medical
device is ready for operation.
Switching on
Prerequisite: The device has been reprocessed
(see chapter ''Cleaning, disinfection and
sterilization'' on page 169) and assembled ready
for operation (see chapter ''Assembly and
preparation'' on page 50).
To prevent condensation and resulting failures of
electrical components, do not switch on the device
after abrupt temperature changes for 1 to 2 hours
(e.g., after storage in unheated rooms).
WARNING
Risk of device malfunction
Some safety systems are only checked
during start-up.
– A selftest should be performed once daily.
– Switch the Fabius on and off or press the
softkey
Run
System
Test
.
WARNING
Risk of explosion and fire
Do not set the device into operation if oxygen
leakage is suspected in the medical device or
its vicinity.
Stop oxygen supply and contact service
personnel.
WARNING
Risk of accidental movement of the medical
device
During operation the medical device can
move accidentally.
Activate the castor brakes.