3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
82
Instrument answer:
[II]B,LLLLLLLLLLUU,YYTTTTTTTTTTUU <CR o CRLF>
in which:
[II]
485 address (only in the case of 485 transmission mode)
B
number of selected channel
,
comma character, ASCII 044
LLLLLLLLLL
gross weight on 10 digits
UU
unit of measure
,
comma character, ASCII 044
YY
2 spaces in the case of null tare or semi automatic tare, PT in the case of preset tare
TTTTTTTTTT tare weight on 10 digits
<CR o CRLF> Terminator, ASCII 13 and ASCII 10 characters
ALIBI MEMORY CANCELLATION (only with non approved instrument)
Command:
[II]ALDL <CR o CRLF>
in which
[II]
485 address (only in the case of 485 transmission mode)
Instrument answer:
[II]ALDLOK <CR o CRLF>
if the cancellation has been carried out effectively
[II]ALDLNO <CR o CRLF>
if the cancellation has not been carried out effectively
5.9 CUSTOMISATION OF THE STRING
[available in ComPC e ComPrn]
Personalise string configuration on the ComPC:
The instrument is able to transmit in a continuous manner or as a response to the READ command on the port linked with
ComPC a data string configurable in the 100 print format through Dinitools™. To restore the transmission of the instrument
strings, one should configure only block "300".
NOTE:
format #100 contains up to 50 macros.
Personalise string configuration on the ComPrn:
The instrument is able to transmit in a continuous manner on the port linked with ComPrn a string of data configurable in
the 01 print format.
For further details in regards to configuring the print formats through Dinitools ™ see the relative manual.