3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
34
See section 6 (
USER MAN.REF.
) for details on the functioning.
(*)
with approved instrument:
- by confirming the setting of ENABLED or CYCLIC it is possible to modify the clearing percentage b/-1
and +/-10 %.
(!) ENABLED, +/-10 %
(*) << 0.PErC >>
ZERO FUNCTIONS IN WEIGHING PHASE
Acquisition of the gross zero through the ZERO key.
Set the clearing percentage in relation to the capacity (in b/-1 and +/- 50 %).
By entering the 0 value, it’s possible to disable the ZERO functions in the weighing phase.
See section 6
(USER MAN.REF.)
for functioning details.
(*)
In case of approved instrument, the settable values are between 0 and 2.
(!) +/-2 %
(*) << 0.trACk >>
ZERO TRACKING
This menu allows setting the zero tracking, in other words, the compensation parameter of the scale’s thermal drift;
the set value corresponds to the number of divisions that is reset in the fixed time of 1 second:
TR. ½
+/- half division.
TR. ¼
+/- one fourth of a division.
TR. 1
+/- one division.
TR. 2
+/- two divisions.
TR. no
tracking disabled.
(!) TR. ½
(*)
with APPROVED instrument it is possible to select just the
TR. no
,
TR. ½
,
TR. ¼
parameters.
(*) << diV.Stb >>
DIVISIONS BY STABILITY
In this step one enters the number of divisions by which the instrument detects the weight stability; the higher the
number of divisions, less is the sensitivity, and consequently the stability is more easily detected. The possible
values are 0…99. By setting the 0 value, the check is disabled.
(!) 02
(*)
with APPROVED instrument, the parameter is read-only.
<< GrAV >>
GRAVITY ZONE AND ZONE OF USE
In this step one selects the gravitational acceleration value
of calibration and of use
of the instrument:
Manual entry of the g value: the instrument is ready for the manual entry of the gravitational acceleration value.
If one enters a wrong g value: the minimum decimal value is proposed (9,75001); by a wrong g value one intends a
decimal number not including between 9,75001 and 9,84999 (included).
(!) g = 9,80655
(*)
with APPROVED instrument the parameter is read only.
(*) << CALib.>>
SCALE CALIBRATION
See section “3.3 SCALE CALIBRATION”.
(*)
with APPROVED instrument the parameters inside of this step are read only.
(*) << 0.CALib. >>
CALIBRATION OF ZERO
See section “3.3 SCALE CALIBRATION”.
(*)
In case of approved instrument the parameter is not displayed.
Clearing percentage
(blinking)
0.PErC
02