3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
36
NO.CTS
Disable
(ITALORA WITTY280 and SMT280)
LOW
CTS active low
(LP522/542, EPSON LX300, TM295, TPR)
HIGH
CTS active high
(DP190)
EMUCTS
Emulation of CTS signal: one is asked to enter the number of characters (nChrS) using 3 digits, which
will be transmitted upon each transmission; then one should enter the wait time in milliseconds (tiME),
using 4 digits, from a transmission and the next one.
XON/XOFF
XON/XOFF control for the printer. Is necessary the insertion of the printer reset command (4
characters in decimal) and the decimal value of XON and XOFF character (17 and 19 of default).
(!) LOW
<< Err.CtS >>
CTS STATUS ERROR
By enabling this error, it is possible to block in advance the print or totalisation function, if recalled with an already
active synchronism signal (see previous step): the indicator display will shows the message “PRINTER ERROR:
CHECK THE CTS!” for a few seconds and return to the weighing phase without carrying out the function.
Press F6/F7 to enable (
enabled
) or disable (
disabled
), and ENTER to confirm.
(!) Disabled
<<
PWrPrn
>>
PRINTER POWER SUPPLY
This step regulates the “AUX” output voltage which is on the board (see the electrical scheme in section 8); one
may select:
PWrEXt
External power supply (AUX output active).
EXtoFF
External auto-off power supply (AUX output always active; at the beginning of each printout some
CR are sent as start-up characters, for a printer in energy saving mode).
PWrint
Internal power supply (AUX output active just when printing).
(!) PWrint
<<
ProtoC
>>
SELECTS PROTOCOL
norMAL
standard protocol
riPE 6
Dini Argeo repeater
Alibi
alibi memory protocol
Cont.
continuous transmission protocol
For the protocol specifications, see section 5.4.
(!) norMAL
<< CoM PC >>
PC SERIAL CONFIGURATION
<<
bAud
>>
SET BAUD RATE
By pressing the
ENTER
key one accesses the selection of the data transmission speed (measured in Baud =
bit/second). The possible values are: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200.
(!) 9600
<<
PAritY
>>
SET PARITY
By pressing the
ENTER
key one accesses the selection of the parity bit type. The possible values are: None
(absent parity bits), Odd (uneven parity bits) e Even (even parity bits).
(!) None
<<
Word
>>
SET WORD
By pressing the
ENTER
key one accesses the selection of the number of data bits. The possible values are: 8 (8
data bits) and 7 (7 data bits).
(!) 8 bit
<<
StoP b
>>
SET STOP BIT
By pressing the
ENTER
key one can then select the number of stop bits. The possible values are: 1 (1 stop bit) and
2 (2 stop bits).
(!) 1 bit
<< CtS.St. >>
SYNCHRONISM SIGNAL
On the serial line set as COM.PRN the indicator can manage a synchronism signal CTS (Clear To Send). A device