3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
49
<< MiLLiV >>
MILLIVOLT
Check of the load cell signal in millivolts, in three decimals.
By pressing
ENTER
the display shows:
If various scales are connected independently by pressing the F6 / F7 keys it is possible to select the scale to be
checked.
By pressing the F8 / F9 keys it is possible to move on the selection of the channel or the DGX (only with DGX type of
cells); the display indicates always the active selection:
By pressing always the F6 / F7 keys it is possible to select the value of the active selection.
In the case the load cell is not connected or faulty, or the A/D converter is faulty, it is possible that floating values are
shown, or the message "Err.C.XX” appears (in which XX is the number of channel / digital load cell
on which the faulty is
detected), if the signal exceed the underload / overload value of the converter.
To check if the A/D converter is faulty, disconnect the channel on which the faulty is detected and make a short between
SIG+ and SIG-; with non faulty A/D converter a mV value proxime to 0.000 will be displayed.
By pressing F5, it is possible to test the minimum value, present value and maximum value of the millivolt
<< ADC POINTS >>
CONVERTER POINTS
Check of the A/D converter points.
By pressing
ENTER
the display shows:
If various scales are connected independently by pressing the F6 / F7 keys it is possible to select the scale to be
checked.
By pressing the F8 / F9 keys it is possible to move on the selection of the channel or the DGX (only with DGX type of
cells); the display indicates always the active selection:
By pressing always the F6 / F7 keys it is possible to select the value of the active selection.
In the case the load cell is not connected or faulty, or the A/D converter is faulty, it is possible that floating values are
shown, or the message "Err.C.XX” appears (in which XX is the number of channel / digital load cell
on which the faulty is
detected), if the signal exceed the underload / overload value of the converter.
Millivolt and type of cells: (A)=Analogue, (D)=Digital
Checked DGX (only with DGX type of cells)
Checked channel / digital load cell (With the DGX type of cells, it is the channel inside the DGX)
Checked scale
MILLIVOLT
0.285 (D)
S1
CH=ALL DGX=ALL
MILLIVOLT
0.285 (D)
S1 CH=ALL
DGX=ALL
A/D converter points
and type of cells: (A)=Analogue, (D)=Digital
Checked DGX (only with DGX type of cells)
Checked channel / digital load cell (With the DGX type of cells, it is the channel inside the DGX)
Checked scale
ADC POINTS
61283 (D)
S1
CH=ALL DGX=ALL
ADC POINTS
61283 (D)
S1 CH=ALL
DGX=ALL