3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
35
(*) << thEo.CA >>
THEORETICAL CALIBRATION
See section “3.3.7 THEORETICAL CALIBRATION”.
(*)
In case of approved instrument the parameter is not displayed.
(*) << AdJ.CEL >>
MANUAL ADJUSTMENT CELLS
See section “3.3.5 ADJUSTMENT OF EQUALIZATION COEFFICIENT”. This parameter is visible only in case of
various cells / channels forming the scale.
(*)
In case of approved instrument the parameter is not displayed.
<< SEriAL >>
SERIALS, PRINTOUTS, ETC…
<< PortS >>
SERIAL CONFIGURATION
By pressing
ENTER
it is possible to select the more suitable combination for the use of the three serial ports in the
indicator hardware (
COM1
,
COM2
,
COM3
):
(!) PC.Pr.AX
<< CoMPrn >>
CONFIGURATION OF PRINTER SERIAL
<< bAud >>
SET BAUD RATE
By pressing the
ENTER
key one accesses the selection of the data transmission speed (measured in Baud =
bit/second). The possible values are: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200.
(!) 9600
<< PAritY
>>
SET PARITY
By pressing the
ENTER
key one accesses the selection of the parity bit type. The possible values are: None
(absent parity bits), Odd (uneven parity bits) e Even (even parity bits).
(!) None
<< Word >>
SET WORD
By pressing the
ENTER
key one accesses the selection of the number of data bits. The possible values are: 8 (8
data bits) and 7 (7 data bits).
(!) 8 bit
<< StoP b >>
SET STOP BIT
By pressing the
ENTER
key one can then select the number of stop bits. The possible values are: 1 (1 stop bit) and
2 (2 stop bits).
(!) 1 bit
<< CtS.St. >>
SYNCHRONISM SIGNAL
On the serial line set as COM.PRN the indicator can manage a synchronism signal.
- by using the dedicated CTS (Clear To Send) signal, if one uses the COM2 port, or
- by using the RX input, if one uses the COM1 or COM3 port (in this case, by enabling the function, this input will be
no longer managed for other reception functions).
A device (like a printer) that is slow in processing the data received, can interrupt the transmission temporarily using
this signal.
With synchronism signal active for a time greater than 10 seconds the indicator cancel the printing, it shows the
message “PRINTER ERROR: CHECK THE CTS!” for a few seconds and returns to the weighing phase.
It is possible to select:
Parameter
COM 1
COM 2
COM 3
PC.Pr.AX
ComPC
ComPrn
ComAux
PC.AX.Pr
ComPC
ComAux
ComPrn
Pr.PC.AX
ComPrn
ComPC
ComAux
Pr.AX.PC
ComPrn
ComAux
ComPC
AX.PC.Pr
ComAux
ComPC
ComPrn
AX.Pr.PC
ComAux
ComPrn
ComPC