3590EKR, 3590EXP, 3590EXT, 3590 EBOX, CPWE, CPWET series indicator E-AF03_05.01_14.07_EN_T.doc
53
From above:
1)
Enter the Setup of the instrument;
(upon start-up, press for an instant the
TARE
key while the instrument version is displayed)
2)
Set the number of scales; enter in the
nuM.SCA
step and select with the
arrow keys:
-
Remote
: only for remote scale (calibrations are not carried out)
-
1 SCALE
-
2 SCALES
and select the type of cells used and the number of digital cells / analogue channels connected for each scale.
1 analogue cell for each scale or until 16 total digital cells.
3)
Enter in the configuration menu of the calibration, in other words
SEtuP >> ConFiG>> CALib.
and press
ENTER
.
4)
Select the scale to be configured with the
keys (only if
nuM.SCA
>
1
) and press
ENTER
.
5)
Execute the equalisation (only in case of more digital cells / channels connected to the scale), like described in section
3.3.5.
6)
Select the calibration menu, in other words
CALib.
and press
ENTER
;
7)
Set the number of calibration decimals: select the
dECiM.
step and press
ENTER
, with the
keys move the
decimal point in the desired position and press
ENTER
.
(!) 0.000
8)
Set the unit of measure: select the
u.M.
step and press
ENTER
, with the
keys select the unit of measure: grams
(- g -), kilograms (- kg -), tons (- t -) or pounds (- lb -) and press
ENTER
.
(!) kg
9)
Select the number of calibration range select the
rAn.nuM
step and press
ENTER
; if there is only one measuring
range, select 1 and confirm with
ENTER;
with various fields (up to 3), the type of functioning will be requested (the
display shows in sequence the message RANGE.T): select “M.Range” (if a multirange scale) or “M.Divis” (if a
multidivisional scale) and confirm with
ENTER
.
10)
Set the division of the scale or the division of the first range: select the
diV. 1
step and press
ENTER
,
with the
keys select the value (1, 2, 5, 10, 20, 50, 100, 200)
and press
ENTER
.
(!) 1
11)
Set the capacity of the scale or the first range: select the
CAP.1
step and press
ENTER
, set the value (minimum 100
maximum 999999) and press
ENTER
.
TAKE NOTE:
enter the whole value including the decimal digits; for example if the capacity should be over 6 kg and
the division 0.001kg (= 1 g), set 6000, or if the capacity should be 1500 kg and the division 0.5 kg, set 15000.
(!) 0.100
NOTE: if the range number
set in step 9)
is equal to 1, pass directly to point
16)
12)
Set the division of the second range: select step
diV. 2
and press
ENTER
, with the
keys select the value (1, 2, 5,
10, 20, 50, 100, 200)
and press
ENTER
.
(!) 1
13)
Set the second range: select step
CAP.2
and press
ENTER
, set the value (minimum 100 maximum 999999) and press
ENTER
.
NOTE: if the range number
set in step 9)
is equal to 2, pass directly to point
16)
14)
Set the division of the third range: select step
diV. 3
and press
ENTER
, with the
keys select the value (1, 2, 5,
10, 20, 50, 100, 200)
and press
ENTER
.
(!) 1