vPad-IV OPERATORS MANUAL
DETAILED TEST DATA:
IV Pump PM Inspection, Channel 1
___________________________________________________________________
Task: Check DUT enclosure for damage
Data: ----
Result: PASS
___________________________________________________________________
Task: Check power cord and plug
Data: ----
Result: PASS
___________________________________________________________________
Task: Check fuse and ratings
Data: ----
Result: PASS
___________________________________________________________________
Task: Check indicators and displays
Data: ----
Result: PASS
___________________________________________________________________
Task: Check battery and battery charger
Data: ----
Result: PASS
___________________________________________________________________
Task: Check air-in-line sensor
Data: ----
Result: PASS
___________________________________________________________________
Task: Run Flow Rate Test
Data: Set Rate (ml/h): 400.0 (tolerance -5 to +5%)
Set VTBI (ml): 20.00 (tolerance -5 to +5%)
Back Pressure (mmHg): 8.3
Duration (hh:mm:ss): 00:03:00
Delivered Volume (ml): 19.89 (error -0.5%)
Average Flow (ml/h): 397.9 (error -0.5%)
Flow Statistics (ml/h):
Mean Max. Min. S.D. 2*SD CV(%)
398.0 445.4 342.0 21.30 42.59 5.35
Result: PASS
___________________________________________________________________
Appendix A
#
Page 80
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...