vPad-IV OPERATORS MANUAL
The priming screen on the tablet (
Figure 17
) summarizes the above instructions and provides a video
which will show you how to connect the input and output tubing and prime the system. After following the
instructions, press the
Check Status
button to confirm the channel has been properly primed. If no air
bubbles are detected in the Module flow sensor, you will then be permitted to set up and run a test. If air
bubbles are detected during the status check, you will be instructed to prime the channel again.
When priming or flushing a channel, always inject
into
the Module and
DO NOT
use the syringe to
withdraw water from the input (
Figure 18
). Excessive or extreme suction applied at the input port of the
Module may result in damage to internal sensors.
Figure 17
Figure 18
Setup/Chapter 3
#
Page 21