vPad-IV OPERATORS MANUAL
primary infusion (or primary infusion to KVO) by direct realtime observation of the flow itself.
Accordingly, the test settings must be configured within the specified limits to ensure the Module will
reliably "see" the change in flow from Rate 1 to Rate 2 when this actually occurs during the test.
Choose a pass/fail
Tolerance
in percent according to specifications of the infusion device or per the
manufacturer's inspection protocol. The tolerance may be applied to Rate 1; VTBI 1; Rate 1 and VTBI 1; or
Rate 1, VTBI 1 and Rate 2 (default).
Once the settings are input, run the test with the
Start
button. With a piggyback or KVO test, always start
vPad-IV
first
, then start the infusion device. Refer to Section 4.5.3 on page 45 for more information on
how the test begins after the
Start
button is pressed.
4.3.6 PCA Pump Test Setup
Use the
PCA Pump Test
to determine the ability of a patient-controlled analgesia (PCA) device to
accurately infuse a set Bolus Volume or dose. The test also measures the lockout time and optional basal
flow between the bolus deliveries. In this test, the PCA device is triggered automatically by vPad-IV to
infuse each bolus during the protocol. Such triggering is accomplished by means of an accessory trigger
cable which connects the vPad-IV Module to the device under test.
Figure 29
Operation/Chapter 4
#
Page 37
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...