vPad-IV OPERATORS MANUAL
4.8.7 More Settings
Press the
More Settings
button of the
Setup
menu (
Figure 43
, page 61) to open the application settings
menu (
Figure 50
).
4.8.7.1 Pressure measurement settings
Use these controls to set the default units of measure for back pressure and occlusion pressure. These are
default settings which are restored each time the app starts up, but which may be changed at test runtime as
required to suit a specific protocol.
As described in Section 4.3.3.6 on page 31, the pressure measurement may be zeroed on each channel
independently. Press the
Pressure Offsets
button to view a menu which displays the current zero offset
for each channel. Via the
Pressure Offsets
menu, zero offsets may be reset individually or collectively.
4.8.7.2 Flow measurement settings
During a test, instantaneous flow is calculated from the incremental increase in volume ∆V which is
observed over a preselected measurement time or "window" ∆T. The instantaneous flow window provides
selections of 10, 30 or 60 seconds for ∆T, with 10 seconds being the default. Higher settings of the window
reduce sensitivity of the instantaneous flow measurement to impulses and other short-term disturbances in
an infusion. A setting of 30 or 60 seconds may be preferable when testing devices such as enteral feeding
pumps, which may have a pulsatile or intermittent delivery characteristic.
Figure 50
Operation/Chapter 4
#
Page 71
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...