vPad-IV OPERATORS MANUAL
Select a pass/fail
Tolerance
in percent from the pick-list according to the manufacturer's inspection
protocol or alternatively per the performance specifications of the device under test. If the test is stopped
when the
Set VTBI
has been delivered, then the tolerance may be applied to the flow rate measured at the
end of the test, or to the volume delivered by the end of the test, or to both the measured rate and the
delivered volume. For other stop conditions, the tolerance applies to flow only since the volume is either
predetermined or the volume is not constrained by the test protocol.
Select the desired units of measure for
Back Pressure
and, if necessary, zero the reading if the pressure
applied to the Module is in fact zero. Refer to Section 4.3.3.6 on page 31 for further details on zeroing.
After zeroing, apply a back pressure to the DUT if this is required by the protocol, following instructions
given in Section 3.5.2 on page 19.
Once the settings are input, run the test with the
Start
button. If the objective is to determine a
volume
as
delivered from the device under test, start vPad-IV
first
, then start the infusion device. If the objective is to
determine accuracy of the
flow
produced by the device under test, you may want to start the infusion device
first, wait a period of time for the flow to stabilize, and then start vPad-IV after the delay. Refer to Section
4.5.3 on page 45 for more information on how the test begins after the
Start
button is pressed.
4.3.4.2 Windowed data analysis
Windowed analysis
of the test results may be optionally enabled in a Flow Rate Test.
Figure 25
is an
example which schematically illustrates the analysis performed on the acquired measurement data after the
Flow Rate Test is completed.
Figure 25
Operation/Chapter 4
#
Page 33
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...