vPad-IV OPERATORS MANUAL
4.5.6 AutoSequence Steps
A vPad-IV AutoSequence will proceed through each pre-programmed step in order, from first step to last.
While an AutoSequence is running, the sequence title and progress (for example, "Step 1 of 5") is shown
just below the channel display. Users can step through the AutoSequence with the control buttons at the
bottom of the screen.
Referring to
Figure 37
, the sequence control buttons are:
Next Step
Proceed to the next step of the AutoSequence
Previous Step
Go back and repeat the previous step of the AutoSequence
Finish Test
End the AutoSequence before all steps are completed. A Test Record may be
saved, but the Record will be assigned a "FAIL". This mode is intended for
creating a record of a failed device inspection in a maintenance system (CMMS),
where the failure could not be addressed at the time of inspection but must wait
until the required maintenance and subsequent re-inspection can be completed at a
later time.
Cancel
Abort the AutoSequence and return to test setup
AutoSequences are intended to follow the steps of an infusion device inspection or maintenance procedure
("PM"). Accordingly, an AutoSequence may include steps which display instructions to the user or which
request a visual inspection of an item or feature of the infusion device. Autsequences may also include steps
which measure flow, volume and occlusion pressure, as described below.
Figure 37
Operation/Chapter 4
#
Page 49
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...