vPad-IV OPERATORS MANUAL
4.3 Test Configuration
4.3.1 Test Setup and Device Recall
Begin a test by selecting an available channel that is displaying a ‘setup and run’ icon
, then enter the
control number or ID of the infusion device to be tested. The device ID, which must have three or more
characters, is used as a primary identifier for Test Records and the on-tablet Equipment Record system.
Enter equipment details you want to be saved in the Test Record by clicking the
Equipment Info
button.
The Equipment Information menu enables you to input manufacturer, model, serial number and similar
details, either by entering it with the on-screen keyboard or with a barcode scanner, or by selecting items
from lists displayed by the
Choose
buttons on the menu. See Section 4.4.1 on page 41 for further
information on the
Equipment Information Editor
.
If the entered device ID is known from a previous test, the app will prompt you to recall the Equipment
Record. Recalling the Equipment Record will load the equipment information and test setup that was saved
for the last test that was performed on the device.
4.3.2 Linking Channels
For testing of infusion devices that contain multiple pumps within a single unit, vPad-IV channels may be
optionally linked during test setup. When the linked channels have completed their respective tests, the
results from all such channels are combined into a single Test Record which is identified by the device ID.
Figure 21
Operation/Chapter 4
#
Page 28
Summary of Contents for vPad-IV
Page 1: ...Operating Manual MN 138a vPad IV Modular Infusion Device Analyzer App Version 1 05 0 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 12: ...vPad IV OPERATORS MANUAL APPENDIX A SAMPLE TEST RECORD 79 Table of Contents Page viii ...
Page 28: ...vPad IV OPERATORS MANUAL Overview Chapter 2 Page 12 ...