AMPS-1 OPERATING MANUAL
CARDIAC OUT INJ
=<
SET TEMP:20°C
Figure 55
- CO INJECTATE
CARDIAC OUT L/M
=<
OUTPUT: 3 L/MIN
Figure 56
- CO L/M
Connect the body temperature cable from the cardiac output monitor to the BODY
connector on the AMPS-1 Cardiac Output adapter.
Connect the injectate temperature cable from the cardiac output monitor to the
INJECTATE connector on the AMPS-1 Cardiac Output adapter.
10.4 Cardiac Output Function (optional)
Note: Cardiac Output Function is an optional feature and must be activated before it
can be used.
Pressing the TEMP/CO button from the Temperature Function (or twice from any
other menu) will change the menu screen to that of
Figure 53
.
Press the UP (
) or DOWN(
) arrow button to scroll through the settings provided
for the baseline body temperature: 36, 37 and 38 °C.
Pressing the RIGHT (
u
) button will change the menu screen to that of
Figure 54
.
Press the UP (
) or DOWN(
) arrow button to toggle between an injectate
temperature of 2 °C and 20 °C.
Pressing the RIGHT (
u
) arrow button will change the menu screen to that of
Figure
55
.
CARDIAC OUT BASE
=<
BASE:37°C
Figure 54
- CARDIAC OUTPUT
Cardiac Output (Optional)/Chapter 10
#
Page 53
Summary of Contents for AMPS-1
Page 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...AMPS 1 OPERATING MANUAL ...
Page 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Page 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Page 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Page 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Page 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Page 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Page 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Page 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Page 100: ......