AMPS-1 OPERATING MANUAL
10.2.4 Cardiac Outputs
• 3, 4, 4.5, 5, 5.5, 6, 6.5 and 7 liters per minute
10.2.5 Simulation Types
• Trend On/Off
• Slow injection
• Faulty injection
• Left-to-right shunt
• Calibration pulse (1 °C for 1 or 4 seconds)
10.3 Setup
Configure the cardiac output monitor to the following settings:
CATHETER TYPE / SIZE:
Baxter Edwards, 93a 131 7f
CALIBRATION COEFFICIENT:
0.542 for 2 °C Injectate
0.595 for 20 °C Injectate
INJECTATE VOLUME:
10 cc
INJECTATE TEMPERATURE:
2 °C or 20 °C
Connect the Cardiac Output Adapter box (P/N 8000-454) to the AUX input on
AMPS-1, using the 8-pin mini-DIN cable supplied with the adaptor.
Figure 53
- Cardiac Output Adapter
Cardiac Output (Optional)/Chapter 10
#
Page 52