AMPS-1 OPERATING MANUAL
ECG BASELINE
=<
IMPEDANCE:1000
Ù
Figure 15
- ECG BASELINE IMPEDANCE
ECG ARTIFACT
=<
ARTIFACT: OFF
Figure 16
- ECG ARTIFACT
Pressing the RIGHT (
u
) arrow button will change the screen to that of
Figure 15
, the
Baseline Impedance Menu.
Press the UP (
) or DOWN(
) arrow button to scroll through the settings provided
for the baseline impedance: 500, 1000, 1500 and 2000
Ù
. These settings simulate the
impedance of the contact between the ECG electrodes and the patient's skin.
Pressing the RIGHT (
u
) an arrow button will change the screen to
Figure 16
, the
ECG Artifact Menu.
Press the UP (
) or DOWN(
) arrow button to scroll through the ECG artifact
simulations: 60Hz, 50Hz, MUSCLE, and WANDERING. "Wandering" artifact uses
the respiration simulation settings for rate and waveshape. The artifact simulation may
be disabled by setting it to OFF (default).
Pressing the RIGHT (
u
) arrow button again will restore the LCD display to the initial
menu screen of
Figure 13
.
Press the LEFT (
t
) arrow button to scroll the display through the above simulator
parameters in reverse order.
Press ENTER after selecting a parameter to activate your selection.
ECG Normal Sinus Rhythm/Chapter 3
#
Page 18
Summary of Contents for AMPS-1
Page 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...AMPS 1 OPERATING MANUAL ...
Page 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Page 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Page 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Page 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Page 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Page 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Page 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Page 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Page 100: ......