AMPS-1 OPERATING MANUAL
8 Pacemaker
8.1 Overview
An artificial pacemaker, or "pacer", is a medical device designed to regulate the beating
of the heart. External pacemakers can be used for initial stabilization of a patient, but
implantation of a permanent pacemaker is sometimes required.
Pacing can be instituted for many reasons: in the case of slow ventricular rate
(bradycardia, heart block) or rapid ventricular rate, atrial fibrillation or atrial flutter, or if
dangerous arrhythmias are noted when the ECG is monitored.
AMPS-1 simulates the presence of a pacer pulse with the ECG signal to allow the
evaluation of how a patient monitor deals with the pacer pulse.
8.2 Available Settings
Press ENTER after selecting a parameter to activate that setting
(Below, * indicates a one-time event, which is activated by pressing the ENTER key)
8.2.1 Rhythms
• Asynchronous at 75 BPM
• Demand with frequent sinus beat (Demand 1)
• Demand with occasional sinus beat (Demand 2)
• A-V sequential
• Non-capture*
• Non-function
8.2.2 Pacer Pulse Amplitudes
• 2, 4, 8, 10, 12, 14, 16, 18, 20, 50, 100, 200, 500 and 700 mV
8.2.3 Pacer Pulse Widths
• 0.1, 0.2, 0.5, 1.0, and 2.0 msec
Pacer/Chapter 8
#
Page 45
Summary of Contents for AMPS-1
Page 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...AMPS 1 OPERATING MANUAL ...
Page 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Page 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Page 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Page 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Page 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Page 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Page 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Page 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Page 100: ......