AMPS-1 OPERATING MANUAL
Figure 69
- AMPSpc Power-On Editor
Note that BP Dynamic settings as listed above do not apply immediately after switching
on the power to AMPS-1. Instead, at power-up AMPS-1 initially sets all BP channels to
zero pressure so that the monitor under test can be zeroed. After zeroing the pressure
channels of the patient monitor, BP Dynamic waveforms as programmed by
AMPSpc
are then activated by pressing
SHIFT-3-ENTER
on the AMPS-1 keypad.
To set or change a power-on default of AMPS-1, click on the "Power On" tab of
AMPSpc
(
Figure 68
).
Double-click on a desired setting in the table (1 - 20), or highlight the setting and press
the "Edit Power On Value" button (pencil icon).
AMPSpc Companion Software/Chapter 13
#
Page 68