AMPS-1 OPERATING MANUAL
AUTO1
=<
NORMAL ADULT
Figure 66
- Auto selection, ALPHA mode
12 Automated Settings
12.1 Overview
The Automated Settings of AMPS-1 provide ten patient simulation setups which are
user-programmable. These setups allow multiple simulation parameters to be
simultaneously modified in as little as two keystrokes, without affecting the power-on
default settings of the simulator.
The Auto setups are created or modified using the
AMPSpc
software, as described in
Section 13.4 of this manual.
12.2 Choosing an Auto Setting
With AMPS-1 in the ALPHA operating mode (the default mode after powering up),
press the AUTO (0) button.
Pressing the LEFT (
t
) or RIGHT (
u
) arrow buttons will scroll the display through the
Auto setups AUTO1 to AUTO10, as indicated in
Figure 65
.
When the desired Auto setup is displayed, press ENTER to activate the parameters of
that Auto.
Auto setups can also be selected when operating AMPS-1 in NUMERIC mode. Press
SHIFT to toggle AMPS-1 from ALPHA to NUMERIC mode, then enter the three
digits corresponding to the desired Auto setting as listed below, then press ENTER to
activate the indicated Auto setup.
Automated Settings/Chapter 12
#
Page 63
Summary of Contents for AMPS-1
Page 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...AMPS 1 OPERATING MANUAL ...
Page 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Page 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Page 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Page 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Page 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Page 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Page 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Page 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Page 100: ......