AMPS-1 OPERATING MANUAL
5 Performance Waveforms
5.1 Overview
AMPS-1 offers a number of performance waveforms to evaluate linearity, frequency
response, R-wave detection and other aspects of the performance of patient monitors
and chart recorders.
When Performance Waves are selected, all BP channels are set to a static pressure of
zero mmHg, and the respiration rate is set to zero.
5.2 Available Settings
Press ENTER after selecting a parameter to activate that setting.
5.2.1 Square Wave
• 2 Hz or 0.125 Hz
5.2.2 Triangle Wave
• 2 Hz
5.2.3 Pulse
• 80 msec, 1 Hz (60BPM)
5.2.4 Sine Waves
• 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
5.2.5 R-Wave Detector Test
• 60 BPM haver-triangle waveform
• Selectable widths: 8, 12, 20, 40, 60, 80, 100, 120, 140, 160, 180 and 200 msec
Performance Waveforms/Chapter 5
#
Page 31