Display:
20 character by 2 line alphanumeric LCD
Dimensions and Weight:
• 10 x 19 x 3 cm (4" x 7
d
" x 1¼")
• 0.34 kg (12 oz)
1.2 Simulation Specifications
ECG General:
• Simulation Type:
Full 12-Lead ECG with independent outputs for each
signal lead, referenced to RL
• Output Impedance:
500, 1000, 1500, and 2000 ohms to RL
• High Level Output:
Lead II amplitude x 500 (0.5V per Lead II millivolt)
• Amplitude Accuracy:
± 2% (Lead II, 2 Hz Square Wave)
Normal Sinus Rhythm:
• Rates:
30, 40, 60, 80, 100, 120, 140, 160, 180, 200,
220, 240, 260, 280, and 300 BPM
• Rate Accuracy:
± 1BPM
• Amplitudes (Lead II):
0.1, 0.25, 0.5, 1, 2, 3, 4 and 5 mV - normal amplitude
settings
0.05 to 5.50 mV in 0.05 mV steps - using performance
settings
• ST Segments:
-0.1, -0.2, -0.3, -0.4, -0.5, -0.6, -0.7, -0.8,
0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7 and 0.8 mV
• Axis Deviation:
Normal (intermediate), horizontal, and vertical.
Modifies ECG baseline during arrhythmias.
• Neonatal Mode:
ECG R-wave width is reduced to 40 milliseconds
Specifications/Chapter 1
#
Page 2
Summary of Contents for AMPS-1
Page 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...AMPS 1 OPERATING MANUAL ...
Page 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Page 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Page 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Page 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Page 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Page 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Page 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Page 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Page 100: ......