AMPS-1 OPERATING MANUAL
TEMPERATURE 400/700
TEMPERATURE 37°C
Figure 52
- TEMPERATURE
9 Temperature
9.1 Overview
AMPS-1 temperature simulations are compatible with Yellow Springs Inc. (YSI) Series
400 and 700 probes. AMPS-1 automatically recognizes and selects the appropriate
simulation based on the adaptor cable connected to the TEMPERATURE jack.
AMPS-1 simulates five body temperatures.
9.2 Available Settings
Press ENTER after selecting a parameter to activate that setting.
9.2.1 Body Temperatures
• 30, 35, 37, 38, and 40 °C
9.3 Temperature Function
In the Temperature function, AMPS-1 automatically senses the adapter connected at
the TEMPERATURE jack (YSI 400 or 700).
Pressing the TEMP/CO(8) button will change the menu screen to that of
Figure 51
.
Temperature/Chapter 9
#
Page 49