6 Specifications and Theory of Operation
1009-0633-000
6-7
Note
Products connected to the electrical outlets may increase the leakage current
above these limits.
w
WARNING
The connection of equipment to the auxiliary mains electrical
outlet(s) may increase the patient leakage currents to values
exceeding the allowable limits in the event of a defective earth
conductor.
Resistance to ground: <0.2 Ω
Power cord
w
WARNING
Use the battery if the integrity of the protective earth conductor is
in doubt.
Battery information
The system is not a portable unit; a sealed lead acid battery supplies battery
backup. Batteries are used as backup power in case of a power failure. Thus,
the battery is in a float charge state most of the time. Batteries meet the
following:
1. Capacity to operate for 90 minutes under typical operating conditions;
30 minutes under extreme conditions.
2. Unit functions to specifications through the transition to battery power.
3. Long float charge life.
4. Battery pack has an auto-resettable thermal fuse.
5. Battery terminals and connecting wires are protected against short
circuits.
Only Datex-Ohmeda service representatives are to replace the battery.
Contact a Datex-Ohmeda service representative to disconnect the battery if
the equipment is not likely to be used for some time. Batteries must be
disposed of in accordance with applicable regulatory requirements in effect at
the time and place of disposal.
Length:
5 meters
Voltage rating:
90 to 264 Vac
Current capacity:
10 A for 220-240 Vac
15 A for 100-120 Vac
Type:
Three conductor power supply cord (medical grade
where required).
Summary of Contents for Aespire 7900