4 Alarms and Troubleshooting
1009-0633-000
4-13
Breathing system problems (no alarm)
Symptom
Problem
Solution(s)
Gas scavenging flow is too low.
Suction supply problem.
Use a different suction supply.
Filter blockage. Active systems
have a flow indicator to show
this.
Replace the filter. Refer to
“Remove the AGSS receiver
filter” in section 2, “
Cleaning and
Strerilization”
.
The bellows fills when the Bag/Vent
switch is set to Bag or the bag fills
when the switch is set to Vent.
Leak through Bag/Vent switch. Ask a qualified service
representative to repair the
system.
The ventilator does not read the
position of the Bag/Vent switch. Use
manual ventilation, if necessary.
Ventilator or absorber
malfunction.
Ventilate manually. Ask a
qualified service representative
to repair the system.
APL valve does not operate correctly. APL valve problem
Replace APL Valve seal and
diaphragm - Refer to User
Maintenance