Aespire 7900
2-24
1009-0633-000
2. Slide the side panel out by removing its tabs from their slots (1).
3. Loosen the thumbscrew and remove the reservoir (2). It is not
autoclavable.
4. Loosen the thumbscrew (1) and lower the receiver to remove it.
5. Replace the filter as necessary. (See following steps.)
6. Do these steps in the opposite order to replace the receiver and the side
panel.
7. Before you use the system, complete the “
Preoperative Tests
” in section
4 of the User Reference Manual, Part 1.
Remove the AGSS receiver filter
The AGSS receiver may be autoclaved. To autoclave AGSS receivers which
have a filter, the filter must be removed because it is not autoclavable.
1
2
1